Checking / Replacing The Particle Filter - TESTO 330-1G Instruction Manual

Flue gas analyzer
Table of Contents

Advertisement

1.800.561.8187
1.800.561.8187
2. Open the condensate outlet on the condensate trap: Pull out
approx. 5 mm against the stop.
3. Let the condensate run out into a sink.
4. Wipe off any drops still on the condensate outlet with a cloth
and close the condensate outlet.
The condensate outlet must be completely closed
(marking), otherwise measuring errors could be caused by
infiltrated air.
7.9.

Checking / replacing the particle filter

Checking the particle filter:
Check the particle filter of the modular flue gas probe periodically
for contamination: check visually by looking through the window
of the filter chamber.
Replace the filter if there are signs of contamination.
Replacing the particle filter:
The filter chamber may contain condensate.
1. Open the filter chamber: Turn gently anti-clockwise.
2. Remove the filter plate and replace it with a new one (0554
3385).
3. Attach the filter chamber and lock it. Turn gently clockwise.