Printing all records for a location:
1. Select measuring location:
2. Start printout:
-
All records for the location are printed out.
Options:
[Options]
>
graphic.
>
[Options]
a record printer.
>
[Options]
>
[Options]
measuring values per display page.
[Options]
>
location.
5.5.

Instrument diagnosis

Important operating values and instrument data are displayed. A
gas path check (testo 330-2 LL) can be carried out. The status of
the sensors and any device errors not yet rectified can be
displayed.
Calling up the function:
>
[
]
Instrument Diagnosis
or
>
[ i
].
Carrying out a gas path check (testo 330-2 LL)
1.
Gas Path Check
2. Place the black sealing cap on the tip of the flue gas probe.
-
The pump flow is displayed. If the volumetric flow rate is less
than 0.02 l/min, the gas paths are not leaking.
3. End of check: [OK].
Viewing device errors:
>
Error Diagnosis
-
Unrectified errors are displayed.
> View next / previous error: [▲], [▼].
www.
www.
[▲], [▼]
[
].
Show
Graphic: Display saved record data as
Print
Data: Transmit data of the chosen record to
Delete
Record: Delete the chosen record.
Number of
Lines: Change the number of
Delete all
Records: Delete all saved records for a
→ [OK].
[OK]
→ [OK].
.com
.com
5 First steps
37
information@itm.com
information@itm.com