Solid Fuel Measurement - TESTO Next-Gen 300 Instruction Manual

Combustion analyzer
Hide thumbs Also See for Next-Gen 300:
Table of Contents

Advertisement

9 Performing the measurement

Solid fuel measurement

9.5.3
For performing solid fuel measurements complying with the 1st
BlmSchV, a Next-Gen testo 300 is required with the following
equipment:
CO sensor H2-compensated for measurements (measuring range for
1st BImSchV measurement min. 20,000 ppm.
Testo 300 Industrial - 30,000 ppm)
Measuring range extension by fresh air dilution
Solid fuel adapter and probe shaft with pre-filter 0600 9765
1
Call up function:
2 |
| Options | Averaging | On
3
Zeroing is carried out.
The list for averaging opens.
The start button changes to
4
Confirm with
5
Enter measuring time (15 min). The required value can be entered in
the field.
6
Confirm with
7
Insert the flue gas probe into the exhaust pipe and position it in the hot
spot.
8
1
Start averaging with
Stability time starts.
Press
After a maximum of 3 minutes, the stabilisation time ends and
the measurement starts automatically.
The system records the readings in the set measuring cycle.
During the measurement, the readings and the calculated
values are displayed.
Readings are automatically saved.
60
The measuring result is displayed after the measuring phase has
been completed.
|
Flue Gas
.
.
.
.
Continue
to end the stability time manually.

Advertisement

Table of Contents
loading

Table of Contents