Caution! Minor physical injury or damage to the equipment may occur. > Apply the specified precautionary measures. Note: Basic or further information. testo 330-1 LL The description only applies for the specified instrument model testo 330-1 LL or testo 330-2 LL. 1..
Use only original spare parts from Testo. > Any further or additional work must only be carried out by authorised personnel. Testo will otherwise refuse to accept responsibility for the proper functioning of the measuring instrument after repair and for the validity of certifications.
> At the end of its useful life, send the product to the separate collection for electric and electronic devices (observe local regulations) or return the product to Testo for disposal. 1.800.561.8187 1.800.561.8187 information@itm.com information@itm.com...
These systems can be adjusted using the testo 330 and checked for compliance with the applicable limit values. The following tasks can also be carried out with the testo 330: • Regulating the O2-, CO- and CO2-, NO-, NOx- values in furnace systems for the purpose of ensuring optimal operation.
The FCC demands that the user is to be informed that with any changes and modifications to the device, which have not been explicitly approved by testo AG, the right of the user to use this device will become null and void.
3 Specifications Parameter Accuracy Response time ΔP ± 0.5 hPa (0.0...50.0 hPa) ±1 % of mv (50.1...100.0 hPa) ±1.5 % of mv (rest of range) Temperature ± 0.5 °C (0.0...100.0 °C) probe dependent ±0.5 % of mv (rest of range) Efficiency Flue gas loss AmbCO2,...
Page 14
3 Specifications Characteristic Values Storage ±0...35 °C temperature battery Battery charge time approx. 5-6 h Battery operation 6h (pump on, 20 °C ambient temperature) time ® Bluetooth (option) Range < 10 m Warranty Measuring instrument: 48 months LL-sensors O2, CO: 48 months NOlow sensor: 12 months Other sensors: 24 months Flue gas probe: 48 months...
Recommended for stowing away the measuring instrument and accessories (example) 4.1.1. Bottom level view 1 Sealing clip 2 testo 330-1 /-2 LL flue gas analyzer 3 Repository for printer accessories • Spare batteries for IRDA printer • 1 roll of spare thermal paper (0554 0568) 4 Repository for printer •...
4 Product description 4.3. Measuring instrument 4.3.1. Overview 1 Switch on/off 2 Interfaces: USB, PS2, infrared CAUTION Risk of injury from infrared beam! > Do not direct infrared beam at human eyes! 3 Condensate trap (on rear) 4 Fixing eyelets for carrying strap (left and right) 5 Display 1.800.561.8187 1.800.561.8187...
Example [▲] Scroll up, increase value [▼] Scroll down, reduce value [esc] Back, cancel function Open main menu [ i ] Open instrument diagnosis menu Transmit data to the Testo protocol printer. 1.800.561.8187 1.800.561.8187 information@itm.com information@itm.com www. www. .com .com...
4 Product description 4.3.3. Display 1 Status bar (dark grey background): • Warning symbol (only if there is a device error, display of error in instrument diagnosis menu), otherwise: Instrument designation. • Symbol (only if data are stored in the temporary memory).
4 Product description 4.3.7. Carrying strap (0440 0581) To secure the carrying strap: > Remove the sealing caps from the sides of the housing. Fix the sealing caps on the inside of the service cover: 1. Place the measuring instrument on its front. 2.
3. Open the battery lock: Press the grey key and push in direction of arrow. 4. Remove the battery and insert a new rechargeable battery. Only use the Testo rechargeable battery 0515 0107! 5. Close the battery lock: Press the grey key and push against direction of arrow until the battery engages.
Switching the measuring instrument on has the effect of stopping battery charging and the measuring instrument is then powered via the mains unit. For longer measurements that are mains-operated, Testo recommends using a combustion air temperature probe with connecting cable. Self-heating of the instrument during mains operation may influence the combustion air temperature measurement with a mini ambient air probe.
5 First steps 5.2.2. Connecting probes / sensors Probe/sensor detection at the flue gas socket is carried out continuously. New probes are recognised automatically. Connect a probe to the probe socket before switching on the measuring instrument or start sensor detection manually after changing the probe: [Options] →...
5 First steps 5.2.3. Switching on > press The start screen is displayed (duration: about 5 s). If the voltage supply was interrupted for a longer period: The menu Date / Time is opened. The pressure sensors are set to zero. There is a device error: The Error Diagnosis is displayed.
5 First steps 5.2.7. Printing / saving data Data are printed out via the function key or the menu Options. Data are saved via the menu Options. The menu Options accessed via the left function key and is available in many different menus.
5 First steps 5.2.10. Switching off Unsaved measuring values will be lost when the flue gas analyser is switched off. > Press Possibly: The pump starts and the sensors are rinsed until the switch-off thresholds (O2 > 20 %, other measurement parameters <...
Page 34
5 First steps Show all 1. Select address: [▲], [▼]. 2. Show details: [Details]. 3. Enable a location: select the location → [OK]. The location is activated. [OK] > Open measurements menu: press again. Search 1. Edit search criteria: [►] →...
5 First steps Create a new measuring location: A location is always created under an address. 1. Select the address in which the location is to be created. [Options] → New/Location → [OK]. 3. Enter values or make settings. 4. Finalise the entry: [Finished]. Other location options: >...
5.5. Instrument diagnosis Important operating values and instrument data are displayed. A gas path check (testo 330-2 LL) can be carried out. The status of the sensors and any device errors not yet rectified can be displayed. Calling up the function: >...
Page 37
5 First steps View sensor diagnosis: 1. > Sensor Diagnosis → [OK]. 2. Select sensor. [▲], [▼]. The status of the sensor is indicated by a lamp. A sensor is able to recover. It is therefore possible that the sensor status indication changes from yellow to green or from red to yellow.
6 Using the product Using the product 6.1. Performing settings 6.1.1. Assigning the right function key The right function key can have a function from the Options menu assigned to it. The menu Options is accessed via the left function key and is available in many different menus.
Page 39
6 Using the product Display Parameter Oxygen Carbon dioxide Carbon monoxide Carbon monoxide undiluted Nitrogen monoxide Nitrogen oxide λ Air ratio amCO Ambient carbon monoxide amCO2 Ambient carbon dioxide O2ref Oxygen reference Edrft external draught (micro pressure probe) E-ΔP external differential pressure (micro pressure probe) ExAir Air ratio...
6 Using the product Options: > [Options] → Number of Lines: Change the number of measuring values per display page. > [Options] → Insert Empty Lines: Insert the empty line before the selected line. > [Options] → Delete Line: Delete the selected line. >...
6 Using the product 6.1.2.4. Date / time Date, time mode and time can be set. Calling up the function: > → Instrument Settings → [OK] → Date/Time → [OK] Setting date/time: 1. Select parameter: [◄], [▲], [▼] → [Edit]. 2.
6 Using the product Show or hide measurement types: 1. Select measurement type: [▲], [▼] 2. Enable / disable measurement type: (enabled), (disabled) 3. Save selection: [Finished]. 6.1.2.2. Printer The headers (lines 1-3) and the footers for the printout can be set. The printer that is used can be activated.
6 Using the product 6.1.2.4. Language The menu language can be set. The number of available languages depends on the activated country version, see Country version, page 44. Calling up the function: > → Instrument Settings → [OK] → Language →...
6 Using the product 6.1.2.6. Password protection The password protection is only valid for functions identified by the following symbol: Password protection can be activated / deactivated, the password can be changed. To deactivate the password protection change the password to 0000 (factory setting).
The sensor protection is activated when the threshold is exceeded. • testo 330-1 LL: Switch-off. • testo 330-2 LL: Dilution, if exceeded again: Switch-off. To deactivate sensor protection the thresholds must be set to 0 ppm. Calling up the function: >...
6.1.3.4. Recalibration / adjustment CO and NO sensors can be recalibrated and adjusted. For recalibration / adjustment Testo recommends the use of the calibration adapter 0554 1205. If obviously unrealistic readings are displayed, the sensors should be checked (calibrated) and, if required, adjusted.
6 Using the product 6.1.4. Fuels The fuel can be selected. The fuel-specific coefficients and limits can be set. Apart from the pre-configured fuels, 10 more customer specific fuels can be configured. Fuel parameter, see In order to maintain the measuring accuracy of the instrument one must choose or configure the correct fuel.
6 Using the product 6.1.5. Programs Five measuring programs for different measurement types can be configured and activated. The measuring programs serve the purpose of saving and representing measuring sequences. After the end of the measuring process the readings of a measuring program are automatically saved in a record.
When the instrument is switched on the measurement menu is opened and the gas sensors are zeroed. testo 330-1 LL: The flue gas probe must be in the open air during the zeroing phase! testo 330-2 LL: The flue gas probe can be in the flue gas duct even during the zeroing phase, if a separate VT- sensor is plugged in.
6 Using the product The tip of the probe must be in the hot spot of the flue gas flow. > Align the flue gas probe in the flue gas duct so that the tip is in the hotspot (area of the highest flue gas temperature). As a visual aid, the display shows the current temperature with a green bar.
6 Using the product 6.2.2. Flue gas To achieve usable measurement results, the measurement period of a flue gas measurement should be approx. 3 min and the measuring instrument should display stable measured values. Calling up the function: → Measurements →...
Page 52
330. Data is transferred via Bluetooth® or via the IrDA interface. For data transfer via Bluetooth®, the testo 315 - 3 and the testo 330 - 2 must have this option, otherwise data is transferred via the IrDA interface.
6 Using the product > [Options] → Show Graphic: The readings are displayed in form of a line graph. [Options] Show Numerical Values: Data are displayed as > → numerical values. > [Options] → System Type: (This function is not available during a measurement).
6 Using the product Options: > [Options] → Clipboard: Data are saved to the clipboard. > [Options] → Delete clipboard: Any data saved to the clipboard is deleted. > [Options] → Save: The readings are saved in a record. [Options] Show Graphic: The readings are displayed in >...
2. Enter data or values → [Next] or [OK]. Determining smoke tester no. / smoke nos. / oil derivative with the smoke tester testo 308 and transferring wireless: The testo 308 must be in data transfer mode ( lights up). [Options] → t308. >...
1. Start measurement: Pressure zeroing. 2. Connect a silicone hose to the testo 330-2 and the system to be tested. The reading is displayed. 3. Quit measurement: Options: >...
6 Using the product Options: [Options] → Clipboard: Data are saved to the clipboard. > > [Options] → Delete clipboard: Any data saved to the clipboard is deleted. > [Options] → Save: The readings are saved in a record. > [Options] →...
6 Using the product 6.2.10. Gas flow The function is only available if the chosen fuel is a gas. Calling up the function: > → Measurements → [OK] → Flow→ [OK]. Performing the measurement: 1. Start measurement: The measuring duration is displayed. 2.
6 Using the product 6.2.12. CO ambient ✓ An ambient CO probe (recommended) or a flue gas probe must be connected. Cigarette smoke influences the measurement by more than 50 ppm. The breath of a smoker influences the measurement by about 5 ppm. When using an ambient CO probe, note that: The direction of flow of the gas has an effect on the accuracy of the measurement.
6 Using the product 6.2.13. CO2 ambient ✓ An ambient CO2 probe (0632 1240) must be connected. In order to obtain correct readings, it is imperative to enter the prevailing absolute pressure. This can be entered Absolute), or it is automatically directly (Pressure calculated when entering...
6 Using the product 6.2.14. Automatic furnaces With the help of the readout adapter for automatic furnaces (0554 1206) status data and error messages can be read out of compatible automatic furnaces, see also documentation on readout adapter. The range of data which can be read out depends on the type of the automatic furnace.
6 Using the product > [Options] → Identification: Information about manufacturer and type of automatic furnace [Options] → Statistics: Display of error statistics. > Automatic furnaces are equipped with a circular buffer memory: Error messages are overwritten when the error log is full.
6 Using the product Options: > [Options] → Clipboard: Data are saved to the clipboard. > [Options] → Delete clipboard: Data in the clipboard are deleted. > [Options] → Save: The readings are saved in a record. [Options] → Show Graphic: >...
6 Using the product 3. Set parameters or enter values: [▲], [▼] and partly [◄], [►]→ [OK]. 4. Pressurise the system. Once the pressure has built up, a stability time specified by DVGW-TRGI 2008 should be observed to ensure that any possible pressure fluctuations are not recorded in the measurement.
6 Using the product 4. [measurement]. Pressure zeroing. 5. Pressurise the system. 6. Start measurement: The stability time will run. After this the measuring phase starts automatically. > End stability time and measurement early: [Next]. The readings and Result Tightness Test 2 are displayed when measurement has been completed.
6 Using the product 6.2.16.4. Leak detection In gas leak detection no measurement is made, but a gas detection is performed. ✓ A gas leak probe (0632 3330) must be connected. You must also refer to the documentation that comes with the gas leak probe.
Transferring data 6.3.1. Report printer To be able to transmit data via infrared or Bluetooth interface to a Testo report printer, the printer to be used must have been activated, see Printer, page 43. Printing out data takes place via [Print] ].
7 Maintaining the product Maintaining the product 7.1. Cleaning the measuring instrument > If the housing of the measuring instrument is dirty, clean it with a damp cloth. Do not use any aggressive cleaning agents or solvents! Mild household cleaning agents and soap suds may be used.
7 Maintaining the product 7.3. Recalibrating / adjusting sensors See Sensor settings, page 45. 7.4. Replacing additional filter The additional filter provides added protection should problems occur with the particle filter in the flue gas probe. The additional filter is very rarely contaminated if the measuring instrument is used normally.
7 Maintaining the product 7.5. Cleaning the modular flue gas probe ✓ Disconnect the flue gas probe from the measuring instrument prior to cleaning. 1. Release the probe catch by pressing the key on the probe handle and remove the probe module. 2.
7 Maintaining the product 7.7. Changing the thermocouple 1. Release the probe catch by pressing the key on the probe handle and remove the probe module. 2. Remove the thermocouple plug-in head from the socket using a screwdriver and pull the thermocouple out of the probe shaft. 3.
7 Maintaining the product 2. Open the condensate outlet on the condensate trap: Pull out approx. 5 mm against the stop. 3. Let the condensate run out into a sink. 4. Wipe off any drops still on the condensate outlet with a cloth and close the condensate outlet.
8 Tips and assistance Tips and assistance 8.1. Questions and answers Question Possible causes / solution Rechargeable battery low > Switch to mains operation. Measuring instrument Batteries / rechargeable batteries switches automatically off empty. or cannot be switch on > Charge rechargeable batteries or switch to mains operation.
8.3. Updating the instrument software you can download the current instrument software (Firmware) for testo 330 (registration required). > Unplug the mains unit and switch off the testo 330. 1. Hold [▲] depressed. 2. Plug in the mains unit, keep [▲]...