Selecting Measurement Mode; Configuring Test Parameters; Setting Frequency - JDS Uniphase CellAdvisor JD700B Series User Manual

Base station analyzer, rf analyzer, signal analyzer
Table of Contents

Advertisement

Chapter 13 Using Mobile WiMAX Signal Analyzer

Selecting measurement mode

Procedure
1
Press the MODE hard key.
2
Press the Signal Analyzer soft key.
3
Press the More (1/2) > Mobile WiMAX soft key. The Spectrum mode is set by default.
4
Press the MEASURE hot key, and then select the measurement mode option from the following
choices:
Spectrum
RF Analysis > Channel Power, Occupied BW, Spectrum Emission Mask, or
Spurious Emissions
Power vs Time (Frame)
Modulation Analysis > Constellation, Spectral Flatness, EVM vs Subcarrier, or
EVM vs. Symbol
Auto Measure
Power Statistics CCDF
OTA > Preamble Scanner, Multipath Profile, Preamble Power Trend, or Route Map

Configuring test parameters

Setting frequency

You can set the frequency with either frequency or channel number. If a frequency to be set matches
to the frequency corresponding to the selected channel standard, the instrument calculates its channel
number and updates the screen with it automatically.
Procedure
To set the center frequency:
1
Press the FREQ/DIST hard key.
2
Toggle the Unit soft key and select Freq.
3
Press the Center Frequency soft key.
4
Enter a value by using the numeric keys. You can also use the rotary knob.
5
Select the unit: GHz, MHz, kHz, or Hz.
6
Optional. To define the amount of frequency increment for the rotary knob, complete the following
steps:
a
Press the Frequency Step soft key.
b
Enter a value by using the numeric keys. You can also use the rotary knob.
c
Press the unit: GHz, MHz, kHz, or Hz.
To set the channel number:
1
Press the FREQ/DIST hard key.
2
Toggle the Unit soft key and select Channel.
3
To select the standard channel, complete the following steps:
a
Press the Channel Std soft key. The standard channel window appears.
See "Appendix C – Band, frequency & channel standard" on page 530 for more information.
b
Highlight the band to be measured by using the rotary knob, the arrow keys, or the Page
Up/Page Down soft keys.
c
Press the Select soft key or the rotary knob to confirm the selection.
414
JD700B Series User's Guide