Chapter 11 Using EV-DO Signal Analyzer
can continue your measurement. The measurement settings can be saved and recalled as a file. You
can also use JDViewer, PC application software to configure a measure setup, save as a file, and load
the file on to the instrument.
Procedure
1
Press the MEASURE SETUP hot key.
2
To set the incremental step for the PN offset, complete the following steps:
a
Press the PN Increment soft key.
b
Enter a value by using the numeric keys.
c
Press the Enter soft key.
3
Optional. Go to SAVE/LOAD > Save, and then select the save option from the choices available
for your measurement mode. See "Using save" on page 32 for more information.
Measurement example
Figure 187 Multipath profile measurement with EV-DO OTA signal analyzer

Code domain power

Channels with high correlation factors are determined to be active channels and are indicated as such
on the display. Once the channels are decoded, the analyzer determines the power of each channel
relative to the total signal power.
This measurement helps to verify that each code channel is operating at its proper level and helps to
identify problems throughout the transmitter design from the coding to the RF section. System
imperfections, such as the non-linearity of amplifiers, will present themselves as an undesired
distribution of power in the code domain.
Setting measure setup
After configuring test parameters as described in "Configuring test parameters" on page 335, you can
continue your measurement. The measurement settings can be saved and recalled as a file. You can
also use JDViewer, PC application software to configure a measure setup, save as a file, and load the
file on to the instrument.
366
JD700B Series User's Guide