Chapter 14 Using Cable and Antenna Analyzer
Reflection/RF Out port or at the end of the connected extension cable.
8
Press the Continue soft key to continue calibration.
The calibration progress bar appears.
After completion, the calibration status on the screen changes to
Quick calibration
The Quick Calibration is useful when you want to measure only cable length in DTF measurements.
Procedure
1
Press the MEASURE SETUP hot key.
2
Press the Calibrate soft key.
3
Connect the OPEN connector of the CAL Kit directly to the Cable and Antenna Analyzer
Reflection/RF Out port or at the end of the connected extension cable.
4
Press the Quick Cal. soft key to turn the quick calibration off.
The calibration status on the screen changes back to the previous state.

2-port calibration

You must perform 2-port calibration for 2-port measurements before making a measurement.
Vector measurement
Figure 237 illustrates the connection diagram for calibration in Vector measurements.
Figure 237 Connection for 2-port calibration in vector measurement
1)
Procedure
1
Press the MEASURE SETUP hot key.
2
Press the Calibrate soft key.
The on-screen instruction for the calibration appears.
3
Connect the OPEN connector of the CAL Kit directly to the Cable and Antenna Analyzer
Reflection/RF Out port or at the end of the connected test cable as illustrated in Figure 237-1).
4
Press the Continue soft key to start calibration.
The calibration progress bar appears.
5
Connect the SHORT connector of the CAL Kit directly to the Cable and Antenna Analyzer
Reflection/RF Out port or at the end of the connected test cable.
6
Press the Continue soft key to continue calibration.
452
2)
JD700B Series User's Guide
CAL
ON.