Code Domain Power - JDS Uniphase CellAdvisor JD700B Series User Manual

Base station analyzer, rf analyzer, signal analyzer
Table of Contents

Advertisement

3
To set the incremental step for the PN offset, complete the following steps:
a
Press the PN Increment soft key.
b
Enter a value by using the numeric keys.
c
Press the Enter soft key.
4
Optional. Go to SAVE/LOAD > Save, and then select the save option from the choices available
for your measurement mode. See "Using save" on page 32 for more information.
Measurement example
Figure 154 Constellation measurement with CDMA signal analyzer
NOTE
You can use the LIMIT hot key to analyze your measurements with the user-definable
limit and Pass/Fail indication. See "Setting limit for modulation tests" on page 317 for
more information.

Code domain power

The Code Domain Power (CDP) measures the distribution of power across the set of code channels,
normalized to the total power. It helps to verify that each code channel is operating at its proper level
and to identify problems throughout the transmitter design from coding to the RF section. System
imperfections such as amplifier non-linearity will present themselves as an undesired distribution of
power in the code domain.
Setting measure setup
After configuring test parameters as described in the "Configuring test parameters" on page 297, you
can continue your measurement. The measurement settings can be saved and recalled as a file. You
can also use JDViewer, PC application software to configure a measure setup, save as a file, and load
the file on to the instrument.
Procedure
1
Press the MEASURE SETUP hot key.
2
To set the Pseudo-Noise (PN) offset, complete the following steps:
JD700B Series User's Guide
Chapter 10 Using CDMA Signal Analyzer
311