Plug in your USB drive that has a map file in .mcf file type created in JDMapCreator.
NOTE
The JDMapCreator converts and resizes a map to fit onto your instrument's display. You
must save your map file (.mcf) into the "SavedMap" folder of your USB drive so that you
can load them onto your instrument.
b
Press SAVE/LOAD hot key, and then select Load > Load Map. See "Using load" on page
34 for more information.
3
Press the MEASURE SETUP hot key.
4
Press the Location Setup soft key.
5
Press the Location soft key.
6
Press the Point 1 soft key, and then do one of the following option:
To automatically get the selected point's positioning information with GPS locked, press the
Get GPS Position.
The instrument displays the latitude and logitude information of the signal received by a
Yagi antenna.
To manually determine positioning for the selected point, complete the following steps:
a
Press the Latitude soft key, and then set the properties: North/South, Degree,
Minute, and Second.
b
Press the Longitude soft key, and then set the properties: East/West, Degree,
Minute, and Second.
7
Press the Azimuth soft key to set an azimuth, and then do one of the following:
Without using a compass, turn the rotary knob to align the straight line on the loaded map
with the direction of the connected Yagi antenna.
The instrument calculates the azimuth and the angle appears on the screen.
Measure an azimuth by using a compass, enter the measured angle, and then press the
Enter soft key.
8
Optional. To save a defined position, press the Save Position soft key and then select one of the
soft keys to which you want to save the position information. You can use the Load Position soft
key to load any saved positions.
9
Press the Point 2 soft key, and then repeat steps 6-7.
10
Press the Point 3 soft key, and then repeat steps 6-7.
11
Press the Display Setup soft key.
12
Press the Display Mode soft key, and then select the display mode: Circum, Inscribed, or
Double.
NOTE
Change the display to different mode options and check possible location of the
interfering signal.