Temperature Measurement Specifications; Table B-2. Temperature Measurement Specifications - Advanced Energy Impac 600 Series User Manual

Pyrometer
Table of Contents

Advertisement

®
Advanced Energy
TEMPERATURE MEASUREMENT
SPECIFICATIONS
Table B‑2. Temperature measurement specifications
Description
Temperature measurement
ranges
Temperature subrange
Spectral ranges
Measurement unit
Signal processing
Load
Switching capacity
Hysteresis
57010228-00C
IN 600:
  −40°C to 700°C (−40°F to 1292°F) (MB 7)
IS 600:
  500°C to 1400°C (932°F to 2552°F) (MB 14)
  650°C to 1800°C (1202°F to 3272°F) (MB 18)
  700°C to 2500°C (1292°F to 4532°F) (MB 25)
IGA 600:
  250°C to 1200°C (482°F to 2192°F) (MB 12)
  370°C to 1800°C (698°F to 3272°F) (MB 18)
  400°C to 2500°C (752°F to 4532°F) (MB 25)
IGA 600/23:
  100°C to 1000°C (212°F to 1832°F) (MB 10)
  150°C to 1800°C (302°F to 3272°F) (MB 18)
IN 600/5:
  100°C to 1300°C (212°F to 2372°F) (MB 13)
  400°C to 2500°C (752°F to 4532°F) (MB 25)
User adjustable, minimum span is 51°C (92°F)
IN 600: 8 μm to 14 μm
IS 600: 0.7 μm to 1.1 μm
IGA 600: 1.45 μm to 1.8 μm
IGA 600/23: 2 μm to 2.6 μm
IN 600/5: 5.14 μm
Celsius or Fahrenheit
Digital
0 Ω to 500 Ω
50 V DC; 0.2 A
Negative hysteresis, 2°C to 20°C (4°F to 36°F) adjustable
(without current or value exceeded = open contact)
Unit Specifications
®
Impac
Series 600 Pyrometer
Specification
B‑2