Periodic Calibration; Inspection And Cleaning; Refurbishment Schedule - Advanced Energy Impac 600 Series User Manual

Pyrometer
Table of Contents

Advertisement

®
Advanced Energy
3. Tighten the screw nuts to secure the new sensor head to the sensor cable.
4. Discover (identify) the newly connected sensor head in one of the following
ways.
◦ If using a converter box with a display, select DIS.
◦ In the Infra600 software, click Discover.
◦ Issue the 1-wire search UPP command as presented in
the Universal Pyrometer Protocols (UPP) Commands"
Once discovered, the sensor head will then be ready for use.
 Important

Periodic Calibration

Advanced Energy recommends checking the Impac pyrometer for proper calibration
every 12 months. Contact AE Technical Support Organization for calibration
recommendations.

Inspection and Cleaning

Contamination and damage to the pyrometer can severely impact the measurement
accuracy. Advanced Energy recommends cleaning the pyrometer lens with isopropyl
alcohol using cotton swabs or wipes, or standard cloths for cleaning glasses or photo
lenses.
 Important
Inspecting the pyrometer lens for scratches or nicks is important for the Impac
pyrometer. Damage to the lens can dramatically affect measurement accuracy.
Using a damaged pyrometer will negatively impact the measurement accuracy. If
the pyrometer is physically damaged, return the pyrometer to Advanced Energy for
repair and recalibration or replacement. See

Refurbishment Schedule

To maintain optimal performance of your Impac unit, AE recommends a 3 – 5 year
refurbishment schedule. Contact AE Technical Support Organization within this
timeframe to determine if any refurbishment is recommended for your unit.
57010228-00C
When using an analog box, the newly connected sensor head is
immediately ready for use.
Installation, Setup, and Operation
®
Impac
Series 600 Pyrometer
"Parameterizing with
on page 3‑13.
"Service, Repairs, and
Upgrades".