Table F‑1. Sensor head UPP commands (Continued)
Parameter
Read
<AAXn>me
temperature
subrange
Set temperature
<AAXn>m1<VAL>
subrange
"1" is numeric.
Response time
<AAXn>ez<VAL>
(t
)
90
57010228-00C
Command
Example: 028A0708 = 650° to 1800°
Reports the currently set temperature
measurement subrange for the sensor head.
The reported value represents the temperature
subrange in the selected temperature scale
(Celsius or Fahrenheit).
Value is 8-character hexadecimal: XXXXYYYY
Example: 044C05DC = 1100° to 1500°
Sets a new temperature measurement subrange
for the sensor head. The set value represents
the new temperature subrange in the selected
temperature scale (Celsius or Fahrenheit).
Value is 8-character hexadecimal: XXXXYYYY
Example: 01A10096015E = 150° to 350° for
the sensor head with address 1 connected to the
converter box with address 1.
Sets the response time for the sensor head.
Valid values for IN 600 and IN 600/5:
UPP Commands
Impac
Description
  YYYY = End of the temperature range
  XXXX = Start of the temperature subrange
  YYYY = End of the temperature subrange
  XXXX = Start of the new temperature
subrange
  YYYY = End of the new temperature
subrange
 Important
The minimum span is 51°C (92°F).
  0 = Intrinsic time constant of the pyrometer
(180 ms)
  1 = 0.5 s
  2 = 1 s
  3 = 2 s
  4 = 5 s
  5 = 10 s
®
Series 600 Pyrometer
F‑2