2. Click Create Cal. File and Save to create a new calibration file based on the
calibration data collected in Step 3.2. Calibration above (Figure 12).
3. In the dialog box, save the new calibration file in appropriate location. The file name
must be Calibration.ini as the Calorimeter will not read a calibration file with any other
name.
NOTE: It is recommended to store the calibration file in its own directory, and name the
directory using the current date and other descriptive information as necessary. The
directory should contain no other calibration.ini files. Save the file as "Calibration.ini"
and not as any other name. The calorimeter will not use files with a different name.
4. Click Send Cal. File and Restart (Figure 12).
Figure 12: HPC-CAL calibration software Calorimeter alignment tab
5. In the dialog box, pick the calibration file saved in Step 3 above.
6. Click OK to load the calibration file.
7. The Calorimeter will restart after loading the calibration file. The alignment is
complete when the Calorimeter restarts and completes the boot process.

Features and Settings

Over Temperature Safety

HPC-CAL software monitors the temperature. If an over temperature condition occurs, HPC-
CAL will either produce a warning or shutdown the AC source, depending on the level of the
overtemperature condition. If shutdown occurs, the test will need restarted.

Adaptive Fan Control

The cooling fan on the 1316 chiller will cycle on/off as needed to maintain the chiller
temperature. Fan cycling is normal operation and should not be cause for alarm.