Appendix F. Upp Commands - Advanced Energy Impac 600 Series User Manual

Pyrometer
Table of Contents

Advertisement

UPP Commands
Table F‑1. Sensor head UPP commands
Parameter
Read measuring
<AAXn>ms
value
Read measuring
<AAXn>ms<VAL>
value (repeated)
Read basic
<AAXn>mb
temperature
range
57010228-00C
Command
Reports the most recent temperature
measurement read by the sensor head.
The pyrometer responds with a 5-digit value in
°C or °F, rounded to one decimal place.
For example, if the most recent temperature read
by the sensor head is 256.3°C, the pyrometer
responds with 02563.
• A response of 77770 indicates that the
• A response of 88880 indicates that the
• A response of 75550 indicates that the
• A response of 74440 indicates that the
Reports a set number of temperature
measurements read by the sensor head.
Valid values are 000 to 999.
For example, if you send the command
01N2ms025, sensor head number 2 connected
to the converter box with address 01 will report
the next 25 temperature measurements.
Reports the basic temperature measurement
range of the sensor head. The reported value
represents the temperature range in the selected
temperature scale (Celsius or Fahrenheit).
Value is 8-character hexadecimal: XXXXYYYY
  XXXX = Start of the temperature range
UPP Commands
®
Impac
Series 600 Pyrometer
Description
sensor head is not ready yet.
measuring temperature is higher than either
the basic temperature measurement range
or set subrange.
sensor head temperature is higher than the
maximum sensor head temperature.
sensor head temperature is lower than the
minimum sensor head temperature.
Appendix