Table F‑1. Sensor head UPP commands (Continued)
Parameter
Emissivity
<AAXn>em<VAL>
Clear time
<AAXn>lz<VAL>
interval of the
"l" is alpha.
maximum value
storage (t
)
cl
External
<AAXn>lx<VAL>
clearing of the
"l" is alpha.
maximum value
storage
Sensor head
<AAXn>sa<VAL>
address
57010228-00C
Command
Valid values for IS 600, IGA 600, and IGA
600/23:
Sets the emissivity of the sensor head
(percentage).
Valid values are 0100 to 1200 (10% to 120%).
Sets the clear time interval of the maximum
value storage of the sensor head.
Valid values are 0 to 8.
Simulates an external reset contact.
Sets the sensor head address.
Valid values are 0 to 8 (0 = no address assigned).
UPP Commands
Impac
Description
  6 = 30 s
  0 = Intrinsic time constant of the unit
(10 ms)
  1 = 0.05 s
  2 = 0.25 s
  3 = 1 s
  4 = 3 s
  5 = 10 s
  0 = Off
  1 = 0.1 s
  2 = 0.25 s
  3 = 0.5 s
  4 = 1 s
  5 = 5 s
  6 = 25 s
  7 = External clearing
  8 = Automatic clearing
 Important
This command can only be used when
the clear time of the maximum value
storage (t
) is set to 7 (external clearing).
cl
®
Series 600 Pyrometer
F‑3