Table F-2. Converter Box And Analog Box Upp Commands - Advanced Energy Impac 600 Series User Manual

Pyrometer
Table of Contents

Advertisement

®
Advanced Energy
Table F‑1. Sensor head UPP commands (Continued)
Parameter
Read internal
<AAXn>gt
sensor head
temperature
Read maximum
<AAXn>tm
internal sensor
head
temperature
Serial number
<AAXn>sn
Reference
<AAXn>bn
number
Restart sensor
<AAXn>re
head
Table F‑2. Converter box and analog box UPP commands
Parameter
Analog output
<AA>as<VAL>
For analog box with
multiple connections,
specify the analog out:
<AA>assX<VAL>
57010228-00C
Command
Reports the internal temperature of the sensor
head.
Response in °C: 4-digit value from 0000 to 1800
Response in °F: 4-digit value from 0320 to 3560
Reports the highest internal temperature the
sensor head has reached since it was powered
on.
Response in °C: 4-digit value from 0000 to 1800
Response in °F: 4-digit value from 0320 to 3560
Reports the unique serial number for the sensor
head.
The pyrometer responds with a 7-digit decimal
value.
Reports the reference number for the sensor
head.
The pyrometer responds with a 6-character
hexadecimal value representing the 7-digit
reference number.
Restarts the sensor head (does not affect any
settings).
Command
Sets the analog output for the converter box.
Valid values are 0 to 4.
UPP Commands
Impac
Description
Description
  0 = 0 mA to 20 mA
  1 = 4 mA to 20 mA
  2 = 0 V to 5 V
  3 = Thermocouple type K
  4 = Thermocouple type J
®
Series 600 Pyrometer
F‑4