Table E-2. Sensor Head Parameters - Advanced Energy Impac 600 Series User Manual

Pyrometer
Table of Contents

Advertisement

®
Advanced Energy
Table E‑1. Converter box parameters (Continued)
Parameter
RS/tw
T/REP ON FAIL
Table E‑2. Sensor head parameters
Parameter
S/ADDR
S/RNG BEG
S/RNG END
S/EMI
S/t90
S/tCL
57010228-00C
Wait time for RS-485
Defines how often a request to a sensor head is repeated (in case
of no response or a wrong answer) before an error message is
shown (timeout)
Sensor head address (0 to 8, 0 = no address assigned)
Sets the beginning of the temperature subrange
 Important
The minimum span is 51°C (92°F).
Sets the end of the temperature subrange
 Important
The minimum span is 51°C (92°F).
Emissivity (10% to 120%)
Exposure time
IN 600 and IN 600/5:
  180 ms , 0.5 s, 1 s, 2 s, 5 s, 10 s, or 30 s
IS 600, IGA 600, and IGA 600/23:
  10 ms , 0.5 s, 1 s, 2 s, 5 s, 10 s, or 30 s
Clear time of the maximum value storage (OFF, AUTO,
EXTERN, 25 s, 5 s, 1 s, 0.25 s, 0.05 s, or 0.01 s)
Pyrometer Parameters (Converter Box with Display)
®
Impac
Series 600 Pyrometer
Description
Description
E‑2

Advertisement

Table of Contents
loading