Advanced Energy Impac 600 Series User Manual page 66

Pyrometer
Table of Contents

Advertisement

®
Advanced Energy
Table F‑2. Converter box and analog box UPP commands (Continued)
Parameter
Sensor head for
<AA>ax<VAL>
analog output
For analog box with
multiple connections,
specify the analog out:
<AA>axsX<VAL>
Address
<AA>ga<VAL>
Baud rate
<AA>br<VAL>
57010228-00C
Command
Sets the sensor head that will be used for analog
output of the converter box.
Valid values are N1 to N8 (sensor head number)
or A1 to A8 (sensor head address).
Sets the address of the converter box. The
pyrometer will automatically reset and begin
working with the new address.
Valid values are 00 to 99.
Sets the baud rate for the converter box (RS-232
and RS-485 interface).
Valid values are 2 to 6 and 8.
UPP Commands
Impac
Description
• 00 to 97 = Valid addresses
• 98 = Global address. Use this address
to send entry commands to all the
connected pyrometers. The pyrometers will
not respond to confirm that a change has
been made.
Example: 98fh1 sets the temperature
display of all the connected pyrometers to
°F.
• 99 = Override address. When only one
pyrometer is connected and the converter
box address is unknown, use this address
to send entry commands to the connected
pyrometer. The pyrometer will respond to
confirm that a change has been made.
  2 = 4800 baud
  3 = 9600 baud
  4 = 19200 baud
  5 = 38400 baud
  6 = 57600 baud
  7 = Not allowed
  8 = 115200 baud
®
Series 600 Pyrometer
F‑5