Advanced Energy

Grounding the Pyrometer

If you are using a converter box, make a suitable earth ground connection using a
customer-supplied ground cable as shown in the following figure. This will prevent
transmission interference.
Figure 3‑2. Ground connection for the converter box

Connecting the Sensor Cable

To meet EMC requirements, use a shielded cable. All AE sensor cables are shielded
and will meet this requirement once properly connected to the terminal block on the
converter box.
 Important
If the sensor cable wires are extended, the shield of the cable also needs to be
extended.
Terminal block for cable #2
Terminal block for cable #1
Figure 3‑3. Connecting the sensor cable
57010228-00C
Feedthrough for cable #2
Cable wires and end sleeves
BROWN
BLACK
BLUE
BROWN
BLACK
BLUE
Feedthrough for cable #1
Installation, Setup, and Operation
®
Impac
Series 600 Pyrometer
Washer
O-ring
Shield
Clamp screw
Sensor cable
3‑3