Adjusting the REXA30 ring
Step 1
Remove the protective film from the surface of the REXA.
Measure the run-out on the REXA ring using a Dial Test Indicator (DTI) to. Use a DTI with low
exertion force to avoid scratching the scale surface. A DTI with a ruby ball stylus is recommended as
a further precaution against scratches.
NOTE: At this stage the ring is not firmly fixed, so to avoid causing the ring to shift position, rotate the
ring slowly and smoothly.
Where the DTI shows the lowest radius reading, gently tap the opposite side of the ring on the edge
using a rubber mallet, until the DTI reading is approximately at the 'mid-point' of the run-out.
Find the new lowest radius reading.
Tap the opposite side of the ring with a rubber mallet until the DTI reading is at the 'mid-point' of the
run-out.
Repeat this process until the run-out of the ring is approximately 30 µm TIR (0.0012 inches).
Step 2
Adjust ring position until the DTI (dial test indicator) reading at these points agrees to 10 µm TIR.
NOTE: 52 mm and 57 mm rings do not have 'slot' fiducials marked.
'Slot' fiducial
44
RESOLUTE™ Functional Safety Installation guide and safety manual Siemens DRIVE-CLiQ encoder system
'Slot' fiducial
Use a DTI with low exertion force to avoid scratching
the scale surface. A DTI with a ruby ball stylus is
recommended as a further precaution against scratches.