Renishaw TONiC FS T40x1 Series Functional Safety Manual

Renishaw TONiC FS T40x1 Series Functional Safety Manual

Angle encoder system

Advertisement

Quick Links

Installation guide
M-6688-9043-02-B
TONiC
FS T40x1 RESM20 angle encoder system
RSLM high accuracy linear encoder
Functional Safety installation guide and safety manual

Advertisement

Table of Contents
loading

Summary of Contents for Renishaw TONiC FS T40x1 Series

  • Page 1 Installation guide M-6688-9043-02-B TONiC FS T40x1 RESM20 angle encoder system ™ RSLM high accuracy linear encoder Functional Safety installation guide and safety manual...
  • Page 2: Table Of Contents

    Contents Product compliance System connection – DOP interface Definitions T40x1 readhead and RESM20 compatibility Information for use Readhead mounting and alignment Functional Safety data declaration System calibration Safety function Diagnostic LEDs Certification Output signals Storage and handling Speed TONiC FS T40x1 readhead installation drawing Electrical connections RESM20 installation drawing (‘A’...
  • Page 3: Product Compliance

    Product compliance Patents Features of Renishaw’s encoder systems and similar products are the subjects of the following patents and This document is an installation guide and safety manual, which details the actions required for the safe patent applications: integration of the TONiC FS (Functional Safety) encoder system, as designated by the T4 prefix in the part number, into a functionally safe system;...
  • Page 4: Definitions

    Definitions Functional Safety data declaration WARNING A hazard with a medium risk of injury if not avoided Product identification TONiC FS (Functional Safety) ESD handling The ESD Susceptibility Symbol consists of a triangle, a reaching hand, and a slash IEC 61508 safety data through the reaching hand. The triangle means ‘Caution’ and the slash through the reaching hand means ‘Don’t touch’.
  • Page 5: Safety Function

    ‘Commissioning test’. In the event of failure the affected parts should be returned to Renishaw for further analysis. Using damaged parts invalidates the Functional The following check must be performed when commissioning the TONiC FS system and after any repair of Safety certification.
  • Page 6: Certification

    Certification EU declaration of conformity TONiC FS encoder system Renishaw plc declares under its sole responsibility that the products identified below are in conformity with all relevant Union legislation. Functional Safety Certificate No. FSC001 Name Description Product Name: TONiC FS encoder system Description: TONiC optical encoder system, that is FS certified SIL2 Part no.:...
  • Page 7: Storage And Handling

    Storage and handling TONiC RESM rings are non-contact optical However, in harsh environments such as encoders that provide good immunity against machine tool applications, protection should be contaminants such as dust, fingerprints and provided to prevent ingress of coolant or oil. light oils. Ring and readhead Ring only Readhead only Acetone Acetone N-heptane Storage Operating Humidity...
  • Page 8: Tonic Fs T40X1 Readhead Installation Drawing

    TONiC FS T40x1 readhead installation drawing (RESM ‘A’ section ring shown) Dimensions and tolerances in mm Optical centreline (incremental and reference mark) ‘A’ section – Offset 1.75 ±0.5 ‘B’ section – Offset 3.25 ±0.5 8.75 * (Yaw tol. ±0.4°) (Roll tol. ±0.5°) 2 off mounting holes M2.5 through, 0.25 0.08...
  • Page 9: Resm20 Installation Drawing ('A' Section)

    RESM20 installation drawing (‘A’ section) Nominal Line Mounting holes Readhead external count (mm) (mm) model Dimensions and tolerances in mm diameter θ (mm) (mm) N holes equally spaced on PCD ØDH Ø3.5 through 52.20 30.04 counterbore top face Ø6 × 3 deep 8 192 30°...
  • Page 10: Resm20 Installation Drawing ('B' Section)

    RESM20 installation drawing (‘B’ section) Dimensions and tolerances in mm Nominal Line Mounting holes Readhead N holes equally spaced on PCD ØDH Ø3.5 through external count (mm) (mm) model diameter θ (mm) (mm) 75.40 55.04 N holes equally spaced on PCD ØDH M3 × 0.5 through 11 840 30°...
  • Page 11: Select A Mounting Option

    Select a mounting option Taper mount method Step 1 Mounting shaft specifications Taper mount Interference fit ‘A’ Section Recommended taper roundness: Diameter (mm) Roundness value (mm TIR) 15° ±0.2° 7 min ≤115 0.025 150 to 225 0.050 ≥300 0.075 ‘B’ Section Allow 2 mm for 417 mm, 489 mm and 550 mm rings only. Not applicable Recommended taper diameter (DT): DO (mm)
  • Page 12 Adjust the screws to reduce the range of radial displacement. When adjusting, identify the screw Step 2 location with the lowest radial displacement and tighten that screw, aiming for the average of the highest and lowest indicator readings. Remove protective film from surface of RESM20. Repeat this process until the DTI readings are within ±5 µm at the screw locations.
  • Page 13: Interference Fit Method

    Interference fit method Step 4 Mounting shaft specifications Insert screws into the remaining mounting holes. DO (mm) DS (mm) 30.033 30.017 37.033 37.017 55.039 55.020 80.045 80.023 Step 5 80.045 80.023 Rotate the RESM20 ring, measuring the radial Remove protective film from surface of RESM20. Diameter Tightening 80.045...
  • Page 14: Tonic Fs Quick-Start Guide

    TONiC FS quick-start guide This section is a quick-start guide to installing a TONiC FS system. More detailed information on installing the system is contained in the following sections of the installation guide. INSTALLATION Ensure scale, readhead optical window and mounting faces are clean and free from obstructions. Plug the readhead cable into the Ti/DOP interface under the cover plate and reassemble interface.
  • Page 15: Cable Connection

    Route the cable away from operating environments that will exceed the EMC limits defined in IEC 61800-5-2 Annex E. Use only Renishaw approved cables between the readhead and the interface. The customer is responsible for verifying the product function where the readhead to interface cable has been re-terminated, including the installation of extension cables.
  • Page 16: System Connection - Dop Interface

    Disconnecting the readhead System connection – DOP interface Remove the cover plate on the interface Approved ESD precautions must be followed at all times during readhead and interface electrical (2 × M2.5 hex head screws). connections. Gently lever the connector PCB (on the The readhead is connected to the DOP interface via a small, rugged PCB connector to allow for easy end of the cable) out of the socket.
  • Page 17: T40X1 Readhead And Resm20 Compatibility

    Ensure readhead part number and RESM size range compatibility at the time of ordering and installation. Reference mark position IN-TRAC reference mark is integrated in the scale, radially aligned with the centre of the mounting hole to the left of the ‘Renishaw’ logo ™ within ±0.5 mm. No external actuators or physical adjustment are required.
  • Page 18: Readhead Mounting And Alignment

    Readhead mounting and alignment Mounting brackets Readhead set-up Ensure that the ring, readhead optical window and mounting face are clean and free from obstructions. The bracket must have a flat mounting surface and should enable conformance to the installation tolerances, allow adjustment to the rideheight of the readhead, and be sufficiently stiff to prevent deflection or vibration NOTES: of the readhead during operation.
  • Page 19: System Calibration

    System calibration Step 2 – Reference mark phasing Move the readhead back and forth over the reference mark until the CAL LED stops flashing and Calibration is an essential operation that completes readhead set-up, with the optimum incremental remains off. The reference mark is now phased. and reference mark signal settings stored in the readhead’s non-volatile memory.
  • Page 20: Diagnostic Leds

    Diagnostic LEDs T40x1 readhead LED DOP interface set-up LED Indication Status Alarm Signal Indication Status output Set-up Green Normal set-up; signal level >70% Incremental Purple Normal set-up; signal level 110% to 135% † Orange Acceptable set-up; signal level 50% to 70% Incremental Poor set-up;...
  • Page 21: Output Signals

    P and Q limit actuators are not suitable for ring encoder (RESM) installation. Limits Pink Open collector Limit switch signal detail is included here for information only. Black Please contact your local Renishaw representative if you require limits on your rotary installation. Set-up – Clear Calibrate –...
  • Page 22: Speed

    Speed Clocked output Maximum speed (m/s) option (MHz) 0004 0020 0040 0100 0200 0400 1000 2000 4000 10KD 20KD 5 µm 1 µm 0.5 µm 0.2 µm 0.1 µm 50 nm 20 nm 10 nm 5 nm 2 nm 1 nm 6.48 3.24 1.62 0.648 0.324 0.162 0.065 0.032 5.40 2.70 1.35 0.540 0.270...
  • Page 23: Electrical Connections

    Electrical connections TONiC FS system grounding and shielding Signal termination Analogue outputs Customer electronics Interface Readhead Inner shield Extension cable 120R Output signals − Outer shield Limit output (Limit output for information only) 5 to 24 V IMPORTANT: The outer shield must be connected to the machine earth (Field Ground). The inner shield must be connected to 0 V at receiving electronics only.
  • Page 24: Output Specifications

    Output specifications Digital output signals (DOP only) Analogue output signals Form – Square wave differential line driver to EIA RS422A (except limits P and Q) Incremental 2 channels A and B in quadrature (90° phase shifted) Incremental 2 channels V † and V differential sinusoids in quadrature (90°...
  • Page 25: General Specifications

    Interface to controller Clocked output option Maximum cable length (MHz) 40 to 50 <40 Analogue Renishaw encoder systems have been designed to the relevant EMC standards, but must be correctly integrated to achieve EMC compliance. In particular, attention to shielding arrangements is essential. TONiC RESM20 Functional Safety installation guide and safety manual Original instructions...
  • Page 26: Tonic Ti Interface Drawing

    TONiC Ti interface drawing TONiC DOP interface drawing Dimensions and tolerances in mm Dimensions and tolerances in mm Ø3.5 Clearance holes for M3 cap screws and star washers or alternative 59.3 33.3 4-40 unc hex 4-40 UNC × 2 stand-off 2 positions Calibration button Output 26-way...
  • Page 27 Renishaw reserves the right to change specifications without notice. RENISHAW and the probe symbol used in the RENISHAW logo are registered trade marks of Renishaw plc in the United Kingdom and other countries. apply innovation and names and designations of other Renishaw products and technologies are trade marks of Renishaw plc or its subsidiaries.

This manual is also suitable for:

Tonic resm20

Table of Contents