List Ftp Settings (List Ftp); List Gain - Scanivalve DSA3217-PTP Series Operation And Service Manual

Pressure scanner digital sensor array
Table of Contents

Advertisement

Section 5: Software

LIST FTP Settings (LIST FTP)

Command
Command Syntax
Arguments
Description
Returns

List Gain

Command
Command Syntax
Arguments
Description
Returns
Example
LIST FTP Variables
Returns all of the FTP settings
LIST FTP
All of the FTP setting followed by the prompt.
LIST FTP
SET USERFTP Scanivalve
SET PASSFTP password
SET PATHFTP C:/FTPDIR
SET IPFTP 10.0.1.222
SET FILEFTP Scan
LIST GAIN
LIST G
None
This list is no longer used in the DSA-PTP series. The command will still generate information,
however be padded with 0's. This temperature conversion was replaced with TEMP.
Lists the 16 temperature gains. These are used to convert temperature counts to degrees
Celsius. This is the "M" term in the characterization equation:
Counts&TempB
Counts&TempB
EC'
EC'
TempM
TempM
SET TEMPM0 <term>
SET TEMPM1 <term>
SET TEMPM2 <term>
::
::
::
SET TEMPM13 <term>
SET TEMPM14 <term>
SET TEMPM15 <term>
To view the temperature gain settings, type: LIST G<CR>
SET TEMPM0 0
SET TEMPM1 0
SET TEMPM2 0
::
::
::
SET TEMPM13 0
SET TEMPM14 0
SET TEMPM15 0
28
28
DSA3200-PTP

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DSA3217-PTP Series and is the answer not in the manual?

Table of Contents