Appendix
The pressure value entered was not numeric
ERROR: Insert's type not valid
ERROR: Insert's type must be M
The insert type must be M
ERROR: LIST M start temp not valid
The start temp entered was less than 0, greater than 69, or not numeric
ERROR: List M start temp above <max temp>
The start temp entered was greater than 69
ERROR: LIST M stop temp not valid
The stop temp entered was less than start temp, not numeric, or greater than 69
ERROR: List M stop temp above <max temp>
The stop temp entered was greater than 69
ERROR: LIST A start temp not valid
The start temp entered was less than 0, greater than 69, or not numeric
ERROR: List A start temp above <max temp>
The start temp entered was greater than 69
ERROR: LIST A stop temp not valid
The stop temp entered was less than start temp, not numeric, or greater than 69
ERROR: List A stop temp above <max temp>
The stop temp entered was greater than 69
ERROR: DELETE start temp value not found
The start temp was not entered or was not numeric
ERROR: DELETE start temp not valid
The start temp entered was less than 0, greater than 69, or not numeric
ERROR: DELETE stop temp value not found
The stop temp was not entered or was not numeric
ERROR: DELETE stop temp not valid
The stop temp entered was less than 0, greater than 69, or not numeric
ERROR: Delete start temp above <max temp>
The start temp entered was greater than 69
ERROR: Delete stop temp above <max temp>
The stop temp entered was greater than 69
ERROR: Max errors exceeded
The error buffer is full - more than 15 errors have been detected
ERROR: No Errors
No errors have been detected - this is the normal response to the error command
Communication Errors
ERROR: Closing slave socket with read error
Fatal Error - The DSA-PTP Module has a hardware problem.
ERROR: Could not spawn Tcp Server
Fatal Error - The DSA-PTP Module has a hardware problem.
ERROR: Could not spawn Udp Id Server
Fatal Error - The DSA-PTP Module has a hardware problem.
ERROR: Could not spawn Udp Server
Fatal Error - The DSA-PTP Module has a hardware problem.
ERROR: Could not create a UDP socket
Fatal Error - The DSA-PTP Module has a hardware problem.
ERROR: Could not bind a UDP socket
Fatal Error - The DSA-PTP Module has a hardware problem.
ERROR: sending UDP bytes
Fatal Error - The DSA-PTP Module has a hardware problem.
ERROR: Alignment error
Fatal Error - The DSA-PTP Module has a hardware problem parsing the data
76
76
DSA3200-PTP