Ay1332X2 External Filter Accessory; Installation Instructions - Thermo Scientific EASYpure RoDi Operating Manual

Ultrapure water purification system
Table of Contents

Advertisement

Section 12
Accessories

Installation Instructions

(continued)
AY1332X2 External
Filter Accessory
Installation Instructions
12-2
EASYpure RoDi
4. Carefully and cleanly cut the EASYpure RODI feedwater tubing such
that one piece will connect the EASYpure RODI to the external
ccessory pump. The other piece will be used to connect the facility
water supply to the "inlet" of the accessory pump. Use 90° fittings as
necessary to help route tubing to and from the external pump
accessory.
5. Attach power cable from back of EASYpure RODI into the side of the
Pump Housing.
6. Open facility water supply valve and check AY1332X1 for leaks.
7. Turn power on to the EASYpure RODI system and operate normally.
This accessory should be used when customer site has feedwater with high
particulate and/or high scaling contaminants. If in doubt contact Technical
Services for possible feedwater testing.
Accessory kit includes:
• Filter housing (1)
• Plumbing adapters (2)
• Pressure gauges (2)
• Filter/Antiscalant assembly (1)
• Wall bracket (1)
• Screw/washer sets for bracket/housing connection (4)
1. Turn power off to the EASYpure RODI system.
2. Close facility water supply valve.
3. Completely and securely assemble AY1332X2 components. See Filter
Connection drawing following. (Note orientation of Filter/Antiscalant
assembly). The use of Teflon® tape on NPT treads is recommended to
aid in a secure leak-proof assembly.
4. Using 2 customer-supplied fasteners, install the assembly on a wall in a
location convenient for filter exchange. Be sure NOT to mount the
assembly over anything that could be damaged or be a safety issue if
water spillage occurs.
5. Install the D13321 inlet tubing connection into the outlet plumbing
adapter.