General Specifications

Section 3
Dimensions
12" W x 19" D x 18 1/8" H (30.5 cm x 48.3 cm x 46.0 cm).
Clearances
Sides - 4" (10.1 cm) minimum to allow air flow
Above - 12" (30.5 cm) minimum for UV lamp replacement
Cartridge replacement requires that the back of the unit be accessible to
open door (total depth, unit + open door, = 34") (86.4 cm).Storage
Reservoir Capacity - Approximately 6.5 liters usable, 7.0 liters total
Electrical Requirements
The EASYpure RODI is equipped with two power cords and
corresponding fuses taped to each power cord to be plugged into a
grounded electrical outlet of the appropriate voltage.
Model D13321:
100-240VAC, voltage fluctuations not to exceed ±10%, 50/60Hz
Feedwater Requirements
Types
1
. . . . . . . . . . . . . . . . . . . . . . . . .Tap (Potable)
Turbidity . . . . . . . . . . . . . . . .1.0 N.T.U. maximum
Pressure Range . .2 bar (30 psig) to 6.9 bar (100 psig)
Temperature Range . . . . . . . . . . . .1-35°C (34-95°F)
Minimum Inlet Flow Requirements . . . . . . . .15 lph
pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-10
TDS . . . . . . . . . . . . . . .≤800 (max. ppm as CaCO3)
Silt Density Index . . . . . . . . . . . . . . . . . . . . . . .<5%
Free Chlorine . . . . . . . . . . . . . . . . . . . . . . . .<3 ppm
Langlier Saturation Index . . . . . . . . . . . . . . . . . . .<1
Iron (total as Fe) . . . . . . . . . . . . . . . . . . . .<0.5 ppm
Silica . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .<30 ppm