Display

Refer to Figure 6-1. In addition to displaying the temperature
compensated (25°C) resistivity in M ohms-cm, the display also indicates
operational modes and error indicators. The following is what can appear
on the display:
Three LEDs illuminate to provide system operational status as follows:
RO System Operating: Green LED ON - RO Operation is ON; LED
OFF - RO Flush or RO Operation is OFF.
Low RO Purity: Red LED ON - % rejection between incoming water
and product water is lower than expected; LED OFF - % reject is
good.
Replace RO Prefilter: Red LED ON - Unit has been in service for 6
months, replace RO prefilter; LED OFF - replace prefilter as needed
and reset timer.
Note When the RO operation is automatically turned on, only RO water
that is above the specified % rejection setting (factory set to 75%) will be
put into the tank. Anytime the RO purity falls below the % rejection
setting the water will be diverted away from the tank and sent to drain.
SbY
Er1
rEC
Er3
IdL
Er4
AFL
Er5
FL5
Er6
FL4
Er7
FL3
Add
FL2
FL1
Section 6
Controls and Normal Operation