Thermo Scientific EASYpure RoDi Operating Manual page 50

Ultrapure water purification system
Table of Contents

Advertisement

RO Membrane
Replacement (cont.)
Thermo Scientific
Product
(blue)
Reject
(yellow)
Feed
(green)
Figure 9-4. RO Membrane Installation
8. Remove the top cover by lifting straight up.
9. Locate the RO membrane and carefully unhook the two springs that
secure it.
10. Remove the old membrane by disconnecting the reject and product
tubing from the top of the membrane and the feed water connection
from the bottom. See page 11 for tubing removal instructions.
11. Remove the new RO membrane (FL1332X2) from its packaging. Note
the location of the FEED, PRODUCT and REJECT connections on
the housing. The correct orientation for the installed membrane will be
vertical with the FEED connection down.
Section 9
Maintenance and Servicing
EASYpure RoDi
9-11