Thermo Scientific EASYpure RoDi Operating Manual page 9

Ultrapure water purification system
Table of Contents

Advertisement

Section 1
Safety Information
1-2
EASYpure RoDi
To avoid personal injury:
1. Do not use in the presence of flammable or combustible materials;
fire or explosion may result. This device contains components which
may ignite such materials.
2. Do not use in the presence of highly corrosive substances such as
bleach or acid baths; fire may result.
3. This device is to be used with water feeds only. Cleaning agents must
be used in compliance with instructions in this manual. Failure to
comply with the above could result in explosion and personal injury.
4. Avoid splashing cleaning solutions on clothing or skin.
5. Ensure all piping connections are tight to avoid chemical leakage.
6. Ensure adequate ventilation.
7. Carefully follow manufacturerʼs safety instructions on labels of
chemical containers and material safety data sheets.
8. Depressurize system prior to opening cartridge access door or
removing top cover.
9. This unit is equipped with an ultraviolet lamp. Ultraviolet radiation is
harmful to the eyes and skin. Do not observe the lamp directly while
it is illuminated.
10. Do not operate unit with door open. Inlet pressure may force RO
prefilter out of position.
11. Refer servicing to qualified personnel.
Thermo Scientific