Reverse Osmosis (Ro) Membrane Installation - Thermo Scientific EASYpure RoDi Operating Manual

Ultrapure water purification system
Table of Contents

Advertisement

UV Lamp Installation
(continued)
Reverse Osmosis (RO)
Membrane Installation
Thermo Scientific
Note If UV lamp is not installed properly an "Er3" message will appear for
15 seconds of every minute.
Caution Do not touch the glass portion of the UV lamp. We recommend
that you wear lint-free gloves when handling the lamp. The glass portion
must be free of fingerprints, perspiration, etc. Even a single fingerprint will
reduce the effectiveness of the lamp. If you accidentally touch the glass
portion of the lamp, clean the lamp with a lint-free cloth; use isopropyl
alcohol if required.
Note The UV lamp contains mercury. If broken or no longer needed, do
not dispose of the UV lamp in the trash. Recycle or dispose of the UV
lamp as hazardous waste.
Warning This unit is equipped with an ultraviolet lamp. Ultraviolet
radiation is harmful to the eyes and skin. Do not observe the lamp directly
when it is illuminated.
6. Carefully insert and hold the UV lamp partially into the UV chamber.
7. Connect the UV lamp to the receptacle in the black plastic cover. Slide
the lamp completely into chamber and replace the black plastic cover
on the UV chamber.
8. If an RO membrane is not also being installed or replaced, reinstall the
EASYpure RODI top cover and latch the cartridge access door closed.
1. Remove the RO membrane from its packaging. Note the location of
the FEED, PRODUCT, and REJECT connections on the housing.
The correct orientation for the installed membrane will be vertical with
the FEED connection down. Refer to Figure 5-6.
2. Remove the Ventgard cap and remove the screws securing the
EASYpure RODI top cover. Remove the cover by lifting straight up.
3. Locate the three unconnected tubes labeled FEED, PRODUCT, and
REJECT.
4. Note the orientation of the membrane housing and install the
unconnected tubes into the corresponding connectors on the
membrane housing. To do so, first wet the tube end with water and
push the tube end firmly into the connector. Start with the FEED
followed by the PRODUCT and finally the REJECT.