UV Lamp Replacement
(continued)
Figure 9-3. UV Lamp Installation
Thermo Scientific
4. Remove the screws securing the EASYpure RODI top cover.
5. Remove the top cover by lifting straight up.
6. Locate the UV oxidation chamber and pull the top black plastic cover
off. Do not pull on the cable.
7. While holding on to the lamp, remove the plug from the lamp.
Dispose of the UV lamp appropriately. The used lamp contains
mercury vapor and should not be disposed of in the trash. Recycle or
dispose of the used lamp as hazardous waste.
8. Remove the replacement lamp from its container. DO NOT TOUCH
THE GLASS PORTION OF THE LAMP. It is recommended that
lint free gloves be worn when handling the lamp. The glass portion
must be free of fingerprints, perspiration, etc. Even a light coating of
perspiration will reduce the effectiveness of the lamp.
9. Clean the lamp with isopropyl alcohol and a lint free cloth.
10. Carefully insert and hold the UV lamp partially into the UV chamber.