DIQ/S 281
Outlet
ba76226e02
repair an IQ sensor.
1
Switch on the maintenance condition for the IQ sensor (see
section 4.8.2).
The display of the sensor in the measured value display flashes.
2
Pull the sensor out of the sample.
3
Perform the calibration in the laboratory, cleaning, maintenance or
repair (removing and replacing) of the sensor (about these topics, see
the component operating manual of the corresponding sensor).
4
Submerse the sensor in the sample again.
5
Wait until the measured value no longer changes.
6
Switch off the maintenance condition for the sensor (see section 4.8.3).
The display of the sensor in the measured value display no longer
flashes.

4.8.1 Maintenance condition of IQ sensors

The following diagram gives you an overview of when an IQ sensor is in the
maintenance condition.
03/2023
Operation
75