Xylem DIQ/S 281 Operating Manual page 133

Universal transmitter for one digital sensor
Table of Contents

Advertisement

DIQ/S 281
Test certificates
Ambient
conditions
Electrical data
DIQ/S 281CR2
(240 V AC/DC
mains version)
Electrical data
DIQ/S 281CR2 /24V
(24 V AC/DC
version)
ba76226e02
cETLus, CE
Temperature
Mounting/installa-
tion/maintenance
Operation
Storage
Relative humidity
Mounting/installa-
tion/maintenance
Yearly average
Dew formation
Site altitude
Power supply
Protective class
Overvoltage
category
Power consumption
Supply
03/2023
+ 5 °C ... + 40 °C (+ 41 ... +104 °F)
- 20 °C ... + 55 °C (- 4 ... + 131 °F)
- 25 °C ... + 65 °C (- 13 ... + 149 °F)
≤ 80 %
≤ 90 %
Possible
Max. 2000 m above sea level
Nominal voltage: 100 ... 240 VAC ± 10 %
Frequency:50/60 Hz according to DIN IEC 60038
Line power connection:2 pin, N and L
Line cross-section of mains connection:
Europe:1.5 ... 4.0 mm
USA:AWG 14 ... 12
Fuse rating on the operator side: max. 16 A
II
II
max. approx. 20 W
Nominal voltage: 24 V AC/DC ± 10 %
protective low voltage
SELV (Safety Extra Low Voltage)
AC frequency:50/60 Hz according to DIN IEC 60038
Connection:2 pin
Line cross-section of connections:
Europe:1.5 ... 4.0 mm
USA:AWG 14 ... 12
Fuse rating on the operator side: max. 16 A
Switch-on current: 1.5 A AC/DC (100 ms)
Technical data