GE
D
5273008-100, R
IRECTION
Table 4-8
Step
Change PRF (Pulse Repetition Frequencies)
17
- (Wall Filter)
18
Trace Sensitivity
19
Time Resolution
20
Spectral Average
21
Power output
22
SV Length
23
Wall Filter
4 - 18
9
EVISION
Doppler Mode Touch Panel Controls
Task
Section 4-3 - General Procedure
LOGIQ C S
Expected Result(s)
Velocity scale determines pulse repetition
frequency. If the sample volume gate range
exceeds single gate PRF capability, the
system automatically switches to high PRF
mode. Multiple gates appear, and HPRF is
indicated on the display.
Adjusts the sensitivity to get more accurate
envelope trace.
Adjusts the resolution in frequency domain.
Optimizes the smoothness of the spectrum.
Different levels can be selected.
Optimizes image quality and allows user to
reduce beam intensity. 10% increments
between 0-100%. Values greater than 0.1 are
displayed.
Sizes the sample volume gate.
Wall Filter insulates the Doppler signal from
excessive noise caused from vessel
movement.
B
S
M
ERIES
ASIC
ERVICE
ANUAL