GE
D
5273008-100, R
IRECTION
3-2-2
Installation Warnings
1.) Since the LOGIQ C Series weighs approximately 75 kg without options, preferably two people
should unpack it. Two people are also preferable for installing any additional bulky items.
2.) There are no operator serviceable components. To prevent shock, do not remove any covers or
panels. Should problems or malfunctions occur, unplug the power cord. Only qualified service
personnel should carry out servicing and troubleshooting.
NOTE:
For information regarding packing labels, refer to LABELS ON PACKAGE.
3.) After being transported, the unit may be very cold or hot. If this is the case, allow the unit to
acclimate before you turn it on. It requires one hour for each 2.5°C increment it's temperature is
below 10°C or above 30°C.
Equipment damage possibility. Turning the system on without acclimation after arriving at site
CAUTION
may cause the system to be damaged.
Table 3-3
°C
60
°F
140 131 122 113 104
8
hrs
9
EVISION
Acclimation Time
55
50
45
40
35
30
95
86
6
4
2
0
0
0
Chapter 3 System Setup
LOGIQ C B
25
20
15
10
5
0
77
68
59
50
41
32
0
0
0
0
2
4
S
S
ASIC
ERIES
ERVICE
-5
-10
-15
-20
-25
-30
23
14
5
-4
-13
-22
6
8
10
12
14
16
M
ANUAL
-35
-40
-31
-40
18
20
3 - 3
Need help?
Do you have a question about the LOGIQ C Series and is the answer not in the manual?
Questions and answers
How to divid it into four quadrants
What is dimmer in ge logic c2 colsole usg machine
Where is the dimmer in the logiq USG machine
The dimmer (light adjustment) is located on the left side button of the monitor on the GE LOGIQ C Series ultrasound machine.
This answer is automatically generated