GE
D
5273008-100, R
IRECTION
4-3-14-3
Restore System Presets and Configurations
The restore procedure overwrites the existing database on the local hard drive. Make sure to insert
CAUTION
the correct CD (or USB memory device).
1.) Insert the Backup/Restore CD (or USB memory device) into the drive.
2.) On the Keyboard, press UTILITY.
3.) On the LCD display, press SYSTEM.
4.) On the LCD display, select BACKUP/RESTORE.
NOTE:
If you are not logged in with administrator privileges, the Operator Login window is displayed.
Log on with administrator privileges.
5.) In the Restore list, select Patient Archive, Report Archive and User Defined Configuration.
6.) In the Media field, select the Backup/Restore CD/DVD (or USB memory device).
7.) Select RESTORE.
The system performs the restore. As it proceeds, status information is displayed on the Backup/Restore
screen.
4 - 28
9
EVISION
Figure 4-19 Backup/Restore Menu