GE LOGIQ C Series Basic Service Manual page 85

Hide thumbs Also See for LOGIQ C Series:
Table of Contents

Advertisement

GE
D
5273008-100, R
IRECTION
4-3-8-3
Doppler Mode OP Panel Controls
Table 4-8
Step
1
Adjust Rejection
2
Adjust Sweep Speed
3
Activate Full Timeline
4
Select Display Format
5
Adjust Frequency
6
Trace Direction
7
Invert
8
Auto Calculation
9
Modify Calcs
10
Trace Method
11
Activate Colorize
12
Activate Gray Map
13
Dynamic Range
14
Adjust Angle Correct
15
Adjust Angle Steer
16
Move Baseline
9
EVISION
Doppler Mode Touch Panel Controls
Task
Chapter 4 Functional Checks
LOGIQ C S
Expected Result(s)
Selects a level below which echoes will not be
amplified (an echo must have a certain
minimum amplitude before it will be
processed).
Changes the speed at which timeline is
swept. Press Sweep Speed up/down to
increase/decrease the value.
Displays only timeline screen. Press the Full
Timescreen to activate.
Display layout can be preset to have B-Mode
and Time-motion side-by-side or over-under.
Enables the adjustment of the probe's
operating frequency. Press Frequency and
select desired value. The selected frequency
is displayed in the status window.
Allows to select different trace direction.
Vertically inverts the spectral trace without
affecting the baseline position. Press invert to
invert the spectral trace. The Plus and Minus
signs on the velocity scale reverse when the
spectrum is inverted.
Enables or disables auto calculation.
Activates the window to modify the auto
calculation items.
Allows to select different trace method.
Colorize the gray scale image to enhance the
eyes' discrimination capability. Press the
Cololize, Trackball to cycle through available
maps and press Set to select.
Displays a map window adjacent to the
image. Move the trackball to select the map.
The image reflects the map as scrolled
through the selections. Press Set to select.
Controls how echo intensities are converted
to shades of gray. Click Dynamic Range to
increase/decrease the value.
Estimates the flow velocity in a direction at
an angle to the Doppler vector by computing
the angle between the Doppler vector and
the flow to be measured.
Slant the Color Flow linear image left or right
to get more information without moving
probes. Click Angle Steer to the left to slant
the linear image.
Adjusts the baseline to accommodate faster
or slower blood flows to eliminate aliasing.
B
S
M
ERIES
ASIC
ERVICE
ANUAL
4 - 17