GE LOGIQ C Series Basic Service Manual page 97

Hide thumbs Also See for LOGIQ C Series:
Table of Contents

Advertisement

GE
D
5273008-100, R
IRECTION
4-3-14-4
Archiving Images
1.) Insert the archive media. To format the archive media, select the Utility button on the
2.) Select Connectivity, then Removable Media.
3.) Format the CD. Verify the format if desired.
4.) Images will be moved from the hard drive by date. Therefore, the best way is to label media by
NOTE:
Images will be moved from the hard drive by date. Therefore, the best way to label media is by
date. When images are moved to the archive media, they will be deleted from the system hard
drive. However, the patient database (backed up earlier) maintains pointers to the location of
the images on the archive media.
5.) Select Backup/Restore, then EZBackup/Move.
6.) Select "Backup File Older Than in Days".
Choose the days
which want to older
than current exam
date.
9
EVISION
Keyboard.
date.
Figure 4-20 Format CD Screen
Figure 4-21 EZBackup/Move
Chapter 4 Functional Checks
LOGIQ C S
B
S
ERIES
ASIC
M
ERVICE
ANUAL
4 - 29