GE
D
5273008-100, R
IRECTION
ELECTRICAL SAFETY
Electrical Test Performed
Outlet (correct ground &wiring config.)
System Ground Continuity
Chassis Source Leakage Current - Probe
Chassis Source Leakage Current - Caster
Chassis Source Leakage Current - CRT
Patient Lead Source Leakage
(Lead to Ground)
Patient Lead Source Leakage
(Lead to Lead)
Patient Lead Source Leakage
(Isolation)
Peripheral 1 Leakage Current
Peripheral 1Ground Continuity
Peripheral 2 Leakage Current
Peripheral 2Ground Continuity
Peripheral 3 Leakage Current
Peripheral 3Ground Continuity
Probe Number
(from previous page)
Probe 1:
Probe 2:
Probe 3:
Probe 4:
Probe 5:
Probe 6:
Probe 7:
Probe 8:
Probe 9:
Final Check. All system covers are in place. System scans with all probes as expected.
Accepted by: ______________________________________________________________________
10-22
9
EVISION
Max Value
Allowed
Max Value
Allowed
Section 10-5 - When There's Too Much Leakage Current...
LOGIQ C S
Value
Measured
OK?
PROBES
Max Value
Measured
OK?
B
S
M
ERIES
ASIC
ERVICE
ANUAL
Comments
Comments