GE LOGIQ C Series Basic Service Manual page 95

Hide thumbs Also See for LOGIQ C Series:
Table of Contents

Advertisement

GE
D
5273008-100, R
IRECTION
4-3-14-2
Backup System Presets and Configurations
NOTE:
Always backup any preset configurations before a software reload. This ensures that if the presets need
to be reloaded, after the software update, they will be the same ones the customer was using prior to
service.
1.) Insert a formatted CD into the drive.
2.) On the Keyboard, press UTILITY.
3.) On the LCD display, press SYSTEM.
4.) On the LCD display, select BACKUP/RESTORE.
NOTE:
If you are not logged in as GE Service or with administrator privileges, the Operator Login
window is displayed. Log on with administrator privileges.
5.) In the Backup list, select Patient Archive, Report Archive and User Defined Configuration.
6.) In the Media field, select CD/DVD (or USB memory device).
7.) Select BACKUP.
The system performs the backup. As it proceeds, status information is displayed on the Backup/Restore
screen.
Check here to
backup
presets and
configurations
9
EVISION
Figure 4-18 Backup/Restore Menu
Chapter 4 Functional Checks