GE
D
5273008-100, R
IRECTION
10-4-8-4
Dale 600 Meter Procedure (cont'd)
5.) Set the meter's "FUNCTION" switch to LEAD TO GROUND position to measure the patient lead to
ground leakage current.
6.) Select and test each ECG lead positions (except ALL) of the LEAD selector, testing each to the
power condition combinations.
Table 10-10 Testing Power Conditions
9
EVISION
ECG Power
Meter's Polarity Switch
ON
ON
ON
ON
OFF
OFF
OFF
OFF
Chapter 10 Care & Maintenance
LOGIQ C S
Meter's Neutral Switch
NORM
NORM
REVERSE
REVERSE
NORM
NORM
REVERSE
REVERSE
B
S
M
ERIES
ASIC
ERVICE
ANUAL
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
10-13