Temperature Unit
It is used to set the unit of temperature. ° C and ° F two
options are provided. After the unit is changed, the set
unit would display at the temperature unit position on the
main interface.
Concentration Unit
It is used to set the unit of measurement results. µg/L or
ppb can be selected. Click "Save" and instrument would
switch to the setting unit.
Measurement Type
It is used to set the measurement type. Si, SiO
and
2
2-
SiO
can be selected. The default setting is SiO
:
3
2
2-
60 µg/L SiO
= 28 µg/L Si = 76 µg/L SiO
2
3
Files and Info
The Files and Info is used to display the Instrument
Model, Serial Number, MIDB Firmware version and
Application version.
Access Control
The Access Control interface can be used to set whether
permission control is required or not. When selecting
permission control, the instrument sets multi- level
permissions and can distribute different permission
passwords to different permission groups according to
the needs of users.
Manual Operation
Orion 8030cX Silica Analyzer provides users with the
following Manual Operation menus: Measurement,
Calibration, Validation, Deep Clean, Clean operation,
Drain Cell, Prime, and Air Flush.
34 | Chapter 5
System Running
8030cX Silica Analyzer