Thermo Scientific Orion 8030cX Instruction Manual page 37

Silica analyzer
Table of Contents

Advertisement

Measurement
Measurement
Click "Start" to run measurements according to the
configured
parameters.
parameters and measurement progress is displayed on
this page. Before measurement, it is recommended to
check the measurement parameters. If any modification is
needed, enter the Setup-Measurement Setup menu to
modify the parameters and enter the Manual Operation-
Measurement page again after saving.
During measurement, the current operation can be
aborted by clicking "Stop". It is not recommended to use
the "Stop" function under this circumstance as pressing
"Stop" will cause the instrument to fully flush the lines and
rinse the flow cell.
Calibration
Click "Start" to implement manual calibration. Before
calibration, it is recommended to check the Calibration
-
Range, Points (Double
Standard Concentrations,
calibration failure response mechanism. Default settings
are recommended to begin with. If these parameters need
to be modified, enter the Setup-Calibration Setup page to
modify the parameters and enter the Manual Operation-
Calibration interface again after saving.
During calibration, the current operation can be stopped
by clicking "Stop", but the user is not recommended to use
the "Stop" function under this circumstance.
As described above in the explanation of the Setup
menus, if calibration points is set to Double (two point
calibration), check to ensure that the calibration standards
8030cX Silica Analyzer
Calibration
Validation
Relevant
measurement
two point calibration or Single),
allowable
deviation
Manual
Operation
Deep
Clean
Clean
being used match the concentration settings for Standard
1 Conc.(Default is 20 µg/L) and Standard 2 Conc. (Default
is 200 µg/L). If calibration points is set to single, Standard 1
Conc. (Default is 20 µg/L) should be set, also Calibration
Background should be set according to the background of
reagent solutions, such as 3 µg/L (or use earlier double
point calibration background - intercept). For the range of
1000~5000
µg/L,
recommended to ensure measurement accuracy. Standard
1 Conc. should be set to 1000 µg/L, Standard 2 Conc.
should be set to 4000 µg/L, and the corresponding standard
bottles should be used.
After the instrument calibration is finished, the instrument
will pop up a window to show calibration results. Click
"Confirm" to complete the instrument calibration.
The calibration results of the instrument will be kept in
calibration records. Users can enter the Log-Calibration
Log page to view the calibration results.
Validation
Validation is used to verify the accuracy of the instrument.
and
A validation solution with known concentration is used for
verification. 20 µg/L silica standard is often used for
validation for ultrapure water. Insert the validation tubing
into the validation solution bottle, then click the "Start"
button for validation.
When
conducting the instrument verification, it is
recommended that the validation solution shall be
purchased from official channels.
If the validation fails (based on the Threshold settings), a
new calibration is recommended. This action
Drain
Air
Prime
Cell
Flush
two–point
calibration
System Running
is
strongly
Chapter 5 | 35