Calibration Setup
Calibration Setup is used to configure calibration
parameters including Calibration Range, Calibration
Calibration Parameters
Range
Mode
Calibration Point
Standard 1 Conc.
Standard 2 Conc.
Calibration Background
Action When Fail
Slope Tolerance
Intercept Tolerance
When
configuring
Calibration
concentration, it is recommended to careful check the
actual concentration of the calibration solutions to ensure
they correspond to the appropriate calibration range.
Accurate measurement results can be obtained only when
the calibration concentration settings are correct and the
correct calibration solutions are used.
During the automatic range switching and automatic
calibration, attention should be paid to the range of
automatic calibration. The default automatic calibration is
for the range 0-1000 µg/L with 20 µg/L and 200 µg/L
standard solutions since this range (0-1000 µg/L) is most
commonly used. The measurement error caused by the
changes to environmental conditions and other conditions
can be avoided through calibration. It is recommended to
8030cX Silica Analyzer
Table 5.2: Definition of Calibration Parameters
Sub-options
The Calibration Range of 0 µg/L~ 1000 µg/L is for low range; instrument is
0–1000 µg/L
calibrated with the absorbance of long pathlength.
The Calibration Range of 1000 µg/L~ 5000 µg/L is for high range;
1000–5000 µg/L
instrument is calibrated with the absorbance of short pathlength.
Manual
The instrument is calibrated once when calibration is started.
The instrument is calibrated automatically with an interval (such as 24
Interval
hours) when Calibration Mode of Interval is selected. The Interval default is
set to 168 hours (1 week).
Double or Single can be chosen. The default is set to Double – two point
/
calibration.
The Silica Standard Concentration for standard 1 as silica is set. The
µg/L
default is set to 20 µg/L. If high range is selected, the default is set to
1000 µg/L.
Available when the Calibration Point is set to Double. The Silica Standard
µg/L
Concentration for standard 2 as silica is set. The default is set to 200 µg/L.
If high range is selected, the default is set to 4000 µg/L.
Available when the Calibration Point is set to Single. The Default is set to
µg/L
0 µg/L.
/
None or Stop can be chosen. The default is set to None.
The value is set to validate whether the current calibration is successful or
%
not. The default is set 30%.
The value is set to validate whether the current calibration is successful or
not. The default is set 30.
Setup
for
standard
Mode, Calibration Points, Standard Concentration(s), Action
When Fails and Calibration Tolerance (Slope and Intercept),
shown in the Table 5.2.
Definition of Parameter
conduct auto-calibration at least once a month. Ideally auto-
calibration should be performed once a week. This setting
can be adjusted under calibration mode: interval frequency.
We recommend using default tolerance settings unless
there is a special case where tolerance needs to be
adjusted.
Validation Setup
Validation
Setup
is
parameters including Validation Mode, Tolerance Unit,
Calibration Tolerance, Standard Conc. and Action When
Fails, shown in the Table 5.3.
System Running Chapter 5 | 29
used to configure
validation