Relay Setup includes the On-Off setting and operating
mode setting. When Relay Setup is " On " the available
Relays Setup Parameters
Relay Event Type
Alarm Setup Parameters
Alarm Type
High threshold
Low threshold
Delay
Event Setup Parameters
Event Type
Pre-run/Post-run Functions: According to different site
applications, when an external equipment is required to
run for a period before the Online Silica Analyzer begins
analysis, the Pre-run function can be activated (e.g. in
case that a water sample pre-treatment equipment is
installed).
Similarly, when an external device needs to be triggered
and started after the analyses of the 8030cX Silica
Analyzer is completed, the Post-run function can be
activated.
8030cX Silica Analyzer
Table 5.4.1: Definition of Relays Setup
Sub-options
Off
Default setting is Off.
Alarm
Relay will be On when the set Alarm type is triggered.
Event
Relay will be On when the set Event type happens.
Table 5.4.2: Definition of Alarm
Sub-options
High
When measurement exceeds High threshold, relay will be On.
Low
When measurement is below Low threshold, relay will be On.
When measurement exceeds High threshold or is below Low threshold,
High + Low
relay will be On.
µg/L
The default is set 5000. It can be configured.
µg/L
The default is set to 0. It can be configured.
Seconds
Delay time to set relay on.
Table 5.4.3: Definition of Event
Sub-options
Measure
Relay will be On when measurement starts.
Calibrate
Relay will be On when calibration starts.
Validate
Relay will be On when validation starts.
Clean
Relay will be On when cleaning starts.
Deep Clean
Relay will be On when deep cleaning starts.
Alarm
Relay will be On when alarms happen.
Alert
Relay will be On when alerts happen.
New Data Ready
Relay will be On when new measurement result is ready.
Lack of Sample
Relay will be On when there is no sample water.
Maintenance
Relay will be On when maintenance mode is ON.
operating modes include Alarm and Event. More details
shown in table 5.4.
Definition of Parameter
Definition of Parameter
Definition of Parameter
Pre-run: go to Setup-Input/Output Setup- Relays-Pre/Post
Run page to set the pre-run command.
Post-run: go to Setup-Input/Output Setup- Relays-Pre/Post
Run page to set post-run commands.
System Running Chapter 5 | 31