Step 3: Insert all tubing (reagent 1, reagent 2, reagent 3,
standard sample, water sample and cleaning solution) into
DI water container.
Step 4: Perform the manual prime procedure twice or more
in succession.
Step 5: Take all tubing (reagent 1, reagent 2, reagent 3,
standard sample, water sample and cleaning solution) out
from the DI water sample container and put them in the
air.
Step 6: Perform the manual prime procedure twice or more
in succession.
Step 7: Unplug the power supply of the instrument to turn
off the instrument.
8030cX Silica Analyzer

Long-term Idle

If the instrument is in idle status for a long time (running
only once a day or even less), in order to prevent the
crystallization of residues or sediment of other particulates
in the instrument tubing and especially in the electronic
rotary valve (ERV), which may block the fluidic path and
cause failure when restarting the instrument. It is highly
recommended to enable on-demand Mode when the
instrument is only running once a day or less. The
OnDemand Mode setting can be selected in the Setup
menu under the Deep Clean option. It is also recommended
in this case to enable the auto-initiation during startup. The
clean function is built into the program to clean after the
instrument is idle over 2 hours to ensure a clean start and
accurate measurement.
Regular Maintenance
Chapter 6 | 41