Table 4.1: Electronic rotary valve (ERV) Port and its Port Connection
Installation Direction of Electronic
Rotary Valve (ERV) and Port
Connection
Note:
Only trained and qualified personnel can perform
reagent-related operations.
Different sample delivery tubing must be correctly
inserted into the corresponding reagent bottles
and standard solution bottles.
8030cX Silica Analyzer
Port No.
Port 1
Port 2
Port 3
Port 4
Port 5
Port 6
Reaction/Measurement
Port 7
Port 8
Port 9
Port 10
Intermediate Port
(CTR)
1
Reagent 1
bottle
2
Reagent 2
bottle
3
Reagent 3
bottle
Validation solution
4
bottle
Deep clean solution
5
bottle
Figure 4.1 Reagents, Standard Samples, and Cleaning
Solutions Installation
Description
Cleaning Solution
Standard Sample 2
Standard Sample 1
Water Sample 1
Reagent 3
cell
Reagent 1
Reagent 2
Water Sample 2
Validation
Common port
Ensure the sample injection tubing reaches the
bottom of the bottle.
After installation, ensure that all vent holes of the
bottles are unblocked.
6
Standard solution 1
bottle
7
Standard solution 2
bottle
8
Clean waste bottle/tank
Chemical waste bottle/
9
tank
10
Sampling tube
Remark
To cleaning solution bottle
To standard sample bottle 2
To standard sample bottle 1
To water sample tube Tee
To reagent bottle 3
To Reaction/Measurement
cell
To reagent bottle 1
To reagent bottle 2
To water sample tube Tee
To validation bottle
To the buffer ring
System Startup Chapter 4 | 21