Thermo Scientific Orion 8030cX Instruction Manual page 19

Silica analyzer
Table of Contents

Advertisement

Connection of Power Cord
The instrument is equipped with power wiring
terminals in the electronic box (see Figure 3.9 and
Table 3.1).
It is strongly suggested that the user shall
install an external power switch or a circuit
breaker (with GFCI protection) for the instrument
to facilitate the operator to control. In addition, the
instrument
must
maintenance
replacement.
The instrument power cord is designed to be a
hard connection and is led in through the threaded
conduit on the right side of the case (see Figure
3.10).
The power cord shall be a certified 3-core wire,
with a minimum of 0.75 mm2 /18 AWG, and the
temperature resistance shall not be less than
75 ° C. The instrument power supply must be
grounded.
Table 3.1: Wiring Mode of AC Power Supply
Wire Terminals
Line (L)
Ground (G)
Neutral (N)
Figure 3.9: Wiring of AC Power Supply
8030cX Silica Analyzer
be
powered
off
operations
such
as
Color of Wire
Brown
Black
Yellow
Green
green
Blue
White
during
reagent
Figure 3.10: Power Cord Access Port
Communication Connection
0/4-20 mA analog signal wiring
The instrument provides two 0/4-20 mA analog signal
outputs through the main control board, which can drive a
load of up to 900 Ω . J4 and J5 are 0/4-20 mA analog signal
wiring terminals (see Figure 3.12), and Table 3.2 provides
the description of corresponding functions.
Signal line
wiring port
Figure 3.12: 0/4-20 mA Current Loop Link
Signal line port connections
are recommended:
Port #1
Port 1: Signal line
Port #2
Port 2: Signal line
Port 3: Air inlet
Port #3
Port 4: USB port
Port #4
Figure 3.11: Signal Wiring
Installation
Chapter 3 | 17