can be performed automatically using the settings in
Setup-Validation Setup menu. In the " Action When Fails "
" Calibrate " .
setting,
select
recommended to validate again.
Deep Clean
When dirty tubing or electronic rotary valve (ERV) is
suspected as a source of measurement error, it is
recommended to run the deep clean method. The special
cleaning solution provided by Thermo Fisher is required
for deep clean. It can remove silica deposit, bacteria, dirt,
and surface fouling. Insert the corresponding tube into the
deep cleaning solution, and then click the "Start" button to
perform deep clean.
If deep clean needs to be stopped, click the "Stop" button,
but this operation is not recommended. Thorough rinsing
of the tubing to remove the deep clean solution is
recommended prior to sample analysis. Therefore, it is
recommended to let the instrument automatically finish the
deep clean method to ensure sufficient rinsing of the
tubing.
After deep clean is finished, it is recommended to calibrate
the instrument with standard solutions. Otherwise, the
instrument may perform poorly.
Clean
The clean function is built-in function as a house-keeping
function reducing any prior residual when instrument is
idle over 2 hours. When dirty reaction/measurement cell
or electronic rotary valve (ERV) may affect measurement
similar as that in Deep Clean, it is recommended to run
clean cycle. The special cleaning solution provided by
Thermo Fisher is required for clean. The clean function is
a gentle cleaning procedure to remove light silica deposit
and surface fouling, and it does not require calibration
afterwards. Make sure that the corresponding tube is
inserted into the cleaning solution, and then click the
"Start" button to carry out clean.
36 | Chapter 5
System Running
After
calibration,
it
If clean needs to be stopped, click the "Stop" button, but
this operation is not recommended. Special cleaning
is
solution are used for clean, which may have certain risks.
Therefore, it is recommended to let the instrument
automatically finish clean.
Drain Cell
Drain Cell is for emptying the reaction/measurement cell,
it is recommended before long-time shut down, delivery or
shipping.
Prime
Prime is for eliminating air in the tubing after replacing
reagents, standard solutions and validation solution. The
default Prime Volume is set to 1.2mL.
Air Flush
Air flush is for emptying the solutions in the tubing, it is
recommended to flush All before long-time shut down,
delivery or shipping.
Maintenance
Before entering the maintenance function, a certain
permission is required.
When entering the maintenance interface and enabling the
maintenance mode, the interface will prompt "Change the
instrument
to
operations such as the Hardware Calibration, Diagnostics
and System Management can only be carried out after
entering the maintenance mode.
It is recommended to enter the maintenance mode only
when the analyzer is in the idle state. If the analyzer is in
the busy state, switching to Maintenance Mode will prompt
waiting for current operation to complete.
8030cX Silica Analyzer
maintenance
mode".
Maintenance