Thermo Scientific Orion 8030cX Instruction Manual page 11

Silica analyzer
Table of Contents

Advertisement

In the cold climate, it is recommended that the silica
analyzer be installed in a non-freezing indoor
environment.
Select the installation location as close as possible to the
water sample source, which can reduce the sample delay
as much as possible. To ensure real- time analysis of
online sample, sample water needs to be pumped into the
flow cell of the silica analyzer at above 50ml/min and at a
pressure over 0.2 Bar (3 psi).
The installation location should be free from irritating or
corrosive gases.
The installation of the instrument shall comply with the
following requirements:
The instrument can be mounted on the wall with the
hanging rack or mounting screws through the four holes
on its back panel for operation. The instrument can be
placed on a table top as well. No matter which installation
method is adopted, it is suggested to fasten the instrument
with screws. After installation, the instrument shall not
move freely in the process of operation.
Ensure that the selected wall on which the instrument is to
be installed can support at least 4 times the weight of the
instrument (about 30 Kg).
Enough space should be reserved on the right side of the
instrument to facilitate door opening and wiring.
Enough space should also be reserved on the left side and
the bottom of the instrument to facilitate tubing connection
and wiring.
The instrument should be installed at a position with
suitable height. It is suggested that the screen shall be
level with the operator's level of eyesight.
Remaining level is required for the installation of the
instrument, and it is recommended to use a levelling tool
to ensure proper installation.
Note:
During installation and operation, when the
instrument door and internal fluid passage
mounting panel are opened, it is necessary to
pay attention to the risk that the instrument may
8030cX Silica Analyzer
topple if it is not properly fastened.
Installation
Chapter 3 | 9

Advertisement

Table of Contents
loading

Table of Contents