16 Installing instruments
88
Planmeca Compact i3
2. Insert the tester into the jack on the back of the instrument console.
Compact i dental unit with hanging-tube instruments
Compact i dental unit with balanced instrument arms
3. Check that the meter indicates within 3 bars above or below 1.
The meter may jump when the tester is inserted. If it does, wait for about
one second until the meter stabilises and then check the reading.
If the reading is 4 or more bars away from 1, the unit will not make an
accurate measurement. Contact Planmeca After Sales.